MPM700
The MPM700 is an Android based Patient Monitor with a 7" touch screen display. It has a built-in web server that makes it an IoT capable patient monitor that can be accessed by a web browser from anywhere.
- EKG (5 leads)
- Blood Pressure
- Pulse/SPO2
- Temperature
- Weight (Bluetooth connected)
- Blood Glucose Reader
- Built in 110-200v power supply
- USB input jack for any USB device
- USB output data jack.
- Electronic Patient Record
Price
USD 699.00 (includes 1 set of adult size accessories)
Specifications
ECG
-
measurement range: ±0.5mVp ~ ±5mVp
-
Heart Rate display range: 20bpm ~ 300bpm
-
Heart Rate display accuracy: ±2% or ±2bpm, whichever is greater.
-
Alarm delay time: <=12s
-
Input noise level for ECG amplifier: <=30uVp-p
-
Input impedance: >= 5M ohm (single ended)
-
CMRR (common mode rejection ratio): >=89dB
-
ECG bandwidth: 0.5~40Hz(monitoring mode), 0.05~75Hz(diagnostic mode)
SPO2
- SpO2 measuring range: 35% ~ 99%
- SpO2 measuring accuracy: Arms <=3% for SpO2 range from 70% to 100%
- Pulse Rate measurement: 30~240bpm with 2% or ±2bpm accuracy
- Update rate: 8 pulse beats
- Low perfusion performance: Accuracy of SpO2 & PR still meets the declared
- specification while the pulse amplitude modulation ratio is as low as 0.4%.
NIBP
- Pneumatic pressure measurement: 0~300mmHg with ±3mmHg accuracy
- Air leakage: <2mmHg/minute @ 300mmHg with 500cc cuff volume
- Systolic pressure range: 40~270mmHg for adult, 40~200mmHg for pediatric
- Diastolic pressure range: 10~210mmHg for adult, 10~150mmHg for pediatric
- MAP range: 20~230mmHg for adult, 20~165mmHg for pediatric
- BP accuracy: Mean error <= ±5mmHg, Standard deviation <= 8mmHg.
- Pulse Rate measurement: 40~240bpm with 2% or ±2bpm accuracy
- Blood pressure measuring time:< 30 sec. for typical adult (80/120mmHg, 75bpm) at manual or auto mode;
- Time interval selection at auto mode: 1, 2, 3, 4, 5, 10, 15, 30, 60, 90 minutes.
- Over pressure protection with limit of 320±10mmHg by dual transducers
Respiration
- Respiratory Rate measurement range: 0~120rpm
- Respiratory Rate measurement accuracy: ±10%
Temperature
- Temperature measurement range: 25.0~45.0 °C
- Temperature measurement accuracy: ±0.2 °C
- response time: <=30 sec.
Blood Sugar
- measurement range from 20-600 ml/dl
- 0.7μL Blood sample requirement
- used strip detection
- response time < 10secs
Body Weight
- measurement range from 5kg-180kg
- BMI Caliculations
- Measurement in Lbs and Kgs.
- Bluetooth enabled
Communication Support:
- 3G/GPRS packet data
- Wifi (802.11 a/b/g)
- USB connectivity to PC/Laptop
Display and Control Device:
- Android Tablet
- Laptop/PC
- Mobile Phone